A Practical Guide to Using a New HPLC Column

Table of Contents

Introduction

Getting a new HPLC column is exciting—but also important. It is a core part of your chromatography system. If you use it the right way from the beginning, it can give you better performance and last much longer. In this guide, we’ll show you how to start using a new HPLC column safely and effectively. 

Choose the Right Column for Your Job

Not all columns are the same. Choose one that fits your method and target compounds.

C18 column: Best for most reversed-phase applications. It is highly hydrophobic.

C8 column: Less hydrophobic than C18. It gives shorter retention times.

Phenyl or CN columns: Useful for aromatic or polar compounds.

Silica column: Used in normal-phase chromatography for very polar compounds.

Column selection depends on your analytes and separation goals—understanding the differences helps ensure optimal results.

Open the Package Carefully

·Unbox gently: Avoid dropping or shaking the package.

·Inspect the column: Check for any visible damage to the box, column body, or end fittings.

·Keep the end plugs on: Do not remove the protective caps until you are ready to connect the column to your system — this prevents the packing material from drying out or becoming contaminated.

·Save the documentation: Retain the datasheet and certificate of analysis for future reference — they often contain important storage solvent and performance information.

Starting carefully sets the tone for longer column life and more reliable results!

Inspect the Column

After unboxing, give your column a quick but careful inspection to catch any issues before installation.

·Check the column body for cracks, dents, or scratches.

·Verify end fittings are secure and undamaged — no bent threads or loose caps.

·Look for leakage signs or residue near the connectors.

·Gently shake (if applicable): there should be no rattling — that could mean loose packing material.

If anything seems off, pause and contact the supplier before proceeding. A quick check now can save you hours of troubleshooting later.

Flush and Condition the Column

Before running any samples, properly flushing and conditioning your new HPLC column is essential for achieving optimal performance and reliable results.

1. Connect the Column Correctly

Start by connecting the column according to the flow direction indicated by the arrow on the column. Ensuring proper orientation is crucial for accurate operation.

2. Start with a Low Flow Rate

Set the flow rate to a low value (around 0.2–0.5 mL/min for analytical columns). This helps to avoid excessive pressure and potential damage to the column, especially when it’s still conditioning.

3. Flush with a Compatible Solvent

Next, flush the column with a compatible solvent, typically the same or similar to the storage solvent specified in the datasheet. This step is necessary to remove any residual storage solvents that could affect your analysis.

4. Gradually Introduce Your Mobile Phase

If your mobile phase consists of a mixture (such as a water/organic solvent ratio), gradually introduce it to the column. Begin with 100% organic solvent (e.g., methanol or acetonitrile) and transition slowly to the working mobile phase composition, such as 60% methanol / 40% water. Avoid abrupt solvent changes to prevent column damage. 

5. Equilibrate the Column

Finally, to fully stabilize the column, pass 10–20 column volumes (CV) of the mobile phase through the column. This process ensures the stationary phase is wetted and removes any remaining impurities, allowing for stable and reliable performance.

Flushing for Different Column Types

The flushing procedure may differ depending on the column type you’re using. Here’s a guideline for the most common types:

·For Reversed-Phase Columns (C18, C8, etc.)

Wash: Start by flushing with 10-20 column volumes of a strong solvent like methanol or acetonitrile to remove storage solvents.

Transition to Starting Mobile Phase: Gradually move to your starting mobile phase. For example, transition from 100% methanol to 80% methanol / 20% water, and then to 60% methanol / 40% water. Avoid sudden changes in solvent composition to prevent column damage.

·For Normal-Phase or HILIC Columns

Initial Flush: Flush the column with a less polar solvent, such as hexane or isopropanol, to remove any residual solvents.

Equilibrate: After this, gradually equilibrate with your starting mobile phase.

Install the Column

Proper installation of your HPLC column is critical to ensure optimal performance and longevity. Here’s a step-by-step guide to installing your column the right way:

1.Inspect the Column Packaging

· Before starting, carefully check the packaging for any damage during shipping. Ensure the column is free from visible defects.

· Keep the column in its original packaging until you are ready to install it to avoid contamination or damage.

2.Prepare the System

· Turn off your HPLC system and make sure the system is properly powered down.

· Clean and prepare the connections, ensuring they are free of any debris or residues from previous runs.

3.Align the Flow Direction

· Identify the flow direction marked on the column. This is usually indicated by an arrow on the column’s body or packaging.

· Align the column in the correct orientation to prevent flow blockages or improper filtration.

4.Connect the Column

· Use the appropriate tubing and fittings for the column’s inlet and outlet ports.

· Tighten the fittings securely, but avoid overtightening to prevent damaging the column or the fittings.

· Ensure the connections are leak-proof by performing a quick leak test before proceeding.

5.Check for Column Dead Volume

· Some columns have specific inlet and outlet fittings designed to minimize dead volume. Make sure to use the correct fittings to avoid introducing unwanted extra space that could affect retention times.

6.Pre-Condition the Column (if needed)

· If the column requires conditioning (depending on its type), flush it with an appropriate solvent at a low flow rate. Follow the manufacturer’s instructions for any specific pre-conditioning steps.

· Equilibrate the column with the mobile phase you plan to use to avoid any inconsistencies in performance.

7.Ensure Proper Column Support

· Make sure the column is supported properly in the system, ensuring it is not under stress or positioned in a way that could lead to mechanical damage.

8.Verify System Performance

· Once installed, run a baseline check with your mobile phase. Monitor for any pressure spikes or unusual behavior, as these may indicate an improper installation or a need for further adjustment.

Do Regular Maintenance

Keep your HPLC column in good shape with simple habits:

· Flush the column after each use to remove residues.

· Store it properly with the recommended solvent when not in use.

· Monitor backpressure to catch early signs of clogging.

· Filter your mobile phase and samples to prevent particle buildup.

· Schedule periodic cleaning to maintain peak performance.

Store the Column Properly

Proper storage ensures the longevity of your HPLC column:

· Store the column vertically to prevent damage to the packing material.

· Use the recommended storage solvent to prevent drying out or contamination.

· Seal the column ends with appropriate caps to avoid exposure to air or contaminants.

· Keep it in a cool, dry place away from direct sunlight or extreme temperatures.

Good storage practices help maintain column performance for future use.

Troubleshooting Tips

Problem Reason Solution
High backpressure Blocked frit or dirty sample Flush, or change guard column
Peak tailing Contamination or void Clean column or check fittings
Shifting peaks Mobile phase not consistent Re-equilibrate with more volume
No peaks Flow direction wrong, or leak Check installation and detector

Conclusion

Starting with a new HPLC column is simple—if you follow the right steps. From careful installation to regular cleaning, every action you take helps protect your column and your data. Whether you’re running pharmaceuticals, food analysis, or research samples, following these guidelines will give you better results and more peace of mind.

Have questions or need help choosing the right HPLC column? Contact us at sales@uhplcs.com —we offer professional advice and a full range of high-quality columns to match your needs.

About uHPLCs

UHPLCs is a leading manufacturer of HPLC columns and consumables for liquid chromatography. The company offers a wide range of products, including empty HPLC columns, solvent filters, guard columns, inline HPLC columns, and PEEK consumables. uHPLCs’ products are used in a variety of applications, including pharmaceutical, biotechnology, environmental, and food safety analysis.

UHPLCs is committed to providing high-quality products and services to its customers. The company has a strong team of engineers and scientists who are dedicated to developing innovative products and solutions. uHPLCs also has a global network of distributors and sales representatives who can provide support to customers around the world.

If you are looking for a reliable supplier of HPLC columns and consumables, uHPLCs is the perfect choice. The company’s products are of the highest quality and its services are unmatched in the industry.

HPLC System Connect Diagram by uHPLCs

Contact

Share:

Uncategorized

What Are the Application of HPLC?

High-Performance Liquid Chromatography (HPLC) has become an indispensable analytical technique in laboratories worldwide. Known for its precision, reproducibility, and ability to handle complex mixtures, HPLC

Read More »
Uncategorized

How Long Does an HPLC Column Last?

HPLC columns are essential for achieving accurate and reproducible chromatographic results. However, they have a finite lifespan. Understanding the typical lifetime of an HPLC column

Read More »

Can’t Get Enough?

Get More Information and Quotation
for Our HPLC Column Hardware Today !

Subscribe for exclusive offers and updates on new arrivals